Assay Method Information | |
| WEE1 IC50 Determination |
Description: | IC50 values of compounds against WEE1 kinase enzyme were determined by LanthaScreen Terbium Labeled TR-FRET assay. Kinase assays were performed in 1× kinase buffer (# PV6135, Invitrogen, Life Technologies Grand Island, N.Y.) where total reaction volume was 10 μl in low-volume 384-well plates (#4511, Corning). Serially diluted compounds (3-fold) were incubated with WEE1 Enzyme (1 nM) (# PR7373A, Invitrogen, Life Technologies Grand Island, N.Y.) for 10 min, following which a mixture of ATP (10 M) (# A1852, Sigma, St-Louis, Mo.) and fluorescent-PolyGT substrate (200 nM) (# PV3610, Invitrogen, Life Technologies Grand Island, N.Y.) was added and incubated in dark at room temperature for 1 h. After 1 h, 10 μl stop solution containing Terbium labeled antibody (4 nM) (# PV3529, Invitrogen, Life Technologies Grand Island, N.Y.) and EDTA (# E5134, Sigma, St-Louis, Mo.) (20 mM) in TR-FRET dilution buffer (# PV3574, Invitrogen, Life Technologies Grand Island, N.Y.) was added. Readings were taken in a Synergy Neo Plate reader (BioTek, Winooski) at single excitation of 340 nm and Dual emission at 495 nm and 520 nm respectively. |
Affinity data for this assay | |
---|---|
If you find an error in this entry please send us an E-mail |